Why are the values for Fractional Shortening (FS) different between analysis in M-Mode and B-Mode?

The long axis LV trace in B-Mode calculates the longitudinal FS (typically between 8-15%) whereas the M-Mode trace calculates the radial FS (35% range). We recommend stating in the presentation of your data if longitudinal or radial FS is shown.